Fluke Calibration P3000 Pneumatic Deadweight Testers

Key Features

  • 0.015% of reading accuracy standard (0.008% optional)
  • 1-30 inHg vacuum (0.03-1 bar vacuum)
  • 3-500 psi (0.2-35 bar) pressure
  • inH2O and mbar ranges available
  • Series 3 non-magnetic stainless steel weights
  • Units can be trimmed to local gravity FOC

Fluke Calibration P3000 Pneumatic Deadweight Testers

This versatile instrument provides calibrations from 1-30 inHg vacuum (0.03-1 bar) and 3-500 psi (0.2-35 bar) in one extremely accurate yet easy to use primary standard. A selector valve converts the unit from pressure to vacuum mode without the need to exchange pistons. Units are manufactured to the highest standards with certified accuracies traceable to International Standards Laboratories such as the National Institute of Standards and Technology (NIST). These piston/cylinder type pneumatic standards provide excellent stability, repeatability and accuracy. An optional built-in hand pump can be added for both pressure and vacuum generation.

Fluke Calibration P3100 Hydraulic Deadweight Testers

Pressure Ranges
Oil operated Ranges to 20,000 psi (1400 bar)
Accuracy
Accuracy1 ±0.015% of reading (±0.008% optional)
1. Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range.
Materials of Construction
Standard weight material Series 3 non-magnetic austenitic stainless steel
Weight density 7.8 g/cm³
Optional fractional weights Solution heat treated aluminum
Weight density 2.7 g/cm³
Piston material Tungsten carbide with nickel binder
Density- 15.0 g/cm³
Cylinder material Hardened martensitic steel
Tungsten carbide (water above 500 psi, 35 bar)
Thermal coefficients of expansion Oil piston/cylinder 16.5 ppm/°C
Water (above 500 psi, 35 bar) 11 ppm/°C
General
Test port adaptors 1/8, 1/4, 3/8 and ½ NPT or BSP
Instrument weight 36 lbs (16 kg)
Instrument size (W x D x H) 17.5 x 12 x 8.5
Mass set weight (typical) 80 lbs (36 kg)
Reservoir volume 150 cc
Screw press displacement 5.5 cc
Pump displacement 4.7 cc per stroke
O-ring seal materials Buna N as standard, Viton and EPDM available
Weight Increments
Minimum standard weight increments Up to 500 psi (35 bar): 1 psi (0.1 bar)
500 to 5000 psi (35 to 350 bar): 10 psi (1 bar)
5000 to 20,000 psi (350 to 1400 bar): 20 psi (2 bar)
Model P3112: 1 psi (0.1 bar)
Optional fractional weights Up to 500 psi (35 bar): 0.1 psi (0.01 bar)
500 to 5000 psi (35 to 350 bar): 1 psi (0.1 bar)
5000 to 20,000 psi (350 to 1400 bar): 2 psi (0.2 bar)
Operating Fluid
Mineral oil Shell spindle oil 22, our reference 55-655, viscosity 40 cs @ 20°C
Options
PressCal software Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments.
Fluids Instruments can be supplied for use with Skydrol or brake fluids. Due to the aggressive nature of these fluids the standard nitrile seals and the acrylic reservoir tube will be replaced by either Viton or EPDM seals (as applicable) and an aluminum reservoir tube. For these or any other specialty fluid applications, please consult the factory.